The Water Itself: Temperature, Conductivity and the 5 K Rule

The Water Itself: Temperature, Conductivity and the 5 K Rule

Writer: admin Time:2026-11-16 09:30 Browse:℃

Why the water has its own rules

An ingress test looks like it is about the enclosure. Half the time it is about the water. Two samples from the same batch, tested in the same rig on the same day, can give opposite verdicts because one tank was at 18 °C and the other at 26 °C, or because the water had been reused since morning. Neither the drawing nor the test report mentions the water, which is exactly why it goes wrong quietly.

IEC 60529 addresses this with a single sentence that is easy to read past: the temperature of the water shall not differ from that of the specimen by more than 5 K. It is not a comfort rule. It is a rule against condensation.

What the 5 K actually prevents

Cool a warm enclosure in cold water and the air inside contracts. Air has to come from somewhere, and it comes in through the seal, carrying water vapour with it. Minutes later the enclosure warms back up, the vapour condenses on the inside of the lid, and the technician opens the box to find droplets sitting exactly where a leak would put them.

The physics is straightforward. A 20 K drop in a 1 litre air volume at 101 kPa shrinks that volume by roughly 7 %, which is 70 mL of air drawn in through whatever path exists. If that air is saturated at 25 °C, it carries about 23 g/m³ of water, so 70 mL brings in around 1.6 mg of water. That is small, but it lands as a film on a cold surface and reads as a failure.

ScenarioWater vs sampleWhat you see afterwardsVerdict risk
Sample soaked at 25 °C, tank at 25 °C0 KNothing on the insideBaseline, trustworthy
Sample at 25 °C, tank at 10 °C15 K colderFine droplets on the lid, no tracksFalse fail, most common
Sample at 10 °C, tank at 25 °C15 K warmerAir expands outward, seal pushed outFalse pass, seal stressed the wrong way
Both left in the room for 24 hWithin 1 KNothingCorrect practice

The practical fix costs one night. Leave the sample in the test room overnight so it reaches room temperature, and keep the tank in the same room. A 24 hour soak at 20 ± 5 °C removes the whole class of argument before the test starts.

Conductivity is not in the standard, but it decides readings

If the pass or fail decision is made with an indicator lamp or a low-voltage signal, the water has to conduct. Tap water in most cities sits between 200 and 800 µS/cm, which is enough for a lamp circuit at 40 V through a series resistor. Deionised water can be below 5 µS/cm, and at that level a genuine bridge between two terminals may not light anything.

The reverse problem is dirt. Reused water accumulates metal particles, flux, and traces of oil from previous samples. Conductivity climbs, and a marginal leakage path that would not have shown in clean water now reads as a failure. Neither direction is recorded anywhere.

  • Change the water when it stops being clear, and record the date on the tank.
  • Measure conductivity once a week if the verdict depends on an indicator lamp.
  • Never test a product that has just come out of a salt mist chamber in the same tank.
  • Filter the return line if the rig recirculates.

What to write in the record

A useful water record has five fields: tank temperature, sample temperature before immersion, conductivity, hours since the last water change, and whether the sample was conditioned in the room overnight. Five lines turn "the results were inconsistent" into a finding.

That is the whole point of the 5 K rule. It is not there to make the setup harder. It is there so that the droplets you find afterwards belong to the product and not to the tank.