The jet tests are defined by a flow rate and a pressure. IPX5 asks for 12.5 L/min through a 6.3 mm nozzle at about 30 kPa. IPX6 asks for 100 L/min through a 12.5 mm nozzle at about 100 kPa. In both cases the pressure that matters is the one at the nozzle, and in both cases the gauge is usually somewhere else.
That single placement decision accounts for more failed round robins than any other part of the rig.
Between the pump and the nozzle there is hose, a valve, a filter, and at least two elbows. Each of them costs pressure, and the loss grows with the square of the flow. At 12.5 L/min a 10 m hose of 12 mm bore loses a couple of kPa, which nobody notices. At 100 L/min the same hose can lose 20 to 30 kPa, which is a third of the IPX6 requirement.
| Configuration | Flow | Gauge at pump | Estimated at nozzle |
|---|---|---|---|
| 1 m hose, 19 mm bore, straight | 100 L/min | 100 kPa | 96 kPa |
| 10 m hose, 12 mm bore, two elbows | 100 L/min | 100 kPa | 72 to 80 kPa |
| 10 m hose, 12 mm bore, dirty filter | 100 L/min | 100 kPa | 60 to 70 kPa |
| Same rig at IPX5 flow | 12.5 L/min | 30 kPa | 28 to 29 kPa |
The pattern is worth remembering: the error is small where the flow is small and large exactly where the standard is most demanding.
The fix is to measure where the standard measures. Put a gauge on a tee within 200 mm of the nozzle, or use a nozzle body with a built-in tap. Then adjust the pump until the near-nozzle gauge reads the required value, and record both readings. The difference between them is the health of your plumbing, and it will drift as filters load.
A flow meter belongs in the same place. Pressure alone cannot confirm 100 L/min: a partly blocked nozzle can hold pressure while passing half the flow, and the sample then gets a gentler test than the certificate claims.
Quote the nozzle pressure and the flow, not the pump setting. A certificate that says "100 kPa" without saying where it was measured cannot be reproduced by anybody, including the lab that wrote it two years later.