IPX5 in IEC 60529 is a 6.3 mm bore nozzle delivering 12.5 L/min at 30 kPa measured at the nozzle, held 2.5 to 3 m from the enclosure for at least 3 minutes. IPX6 is the same arrangement with a 12.5 mm bore, 100 L/min and 100 kPa. Those six numbers are the whole specification of the jet. Everything else, the shape of the spray cone, the whiteness of the water, the noise it makes on the sample, is a consequence and not a requirement.
Which is why a stream that looks convincing proves nothing. A nozzle bored out by two hundredths of a millimetre still throws an impressive jet, and an operator watching it has no way to know that the delivery rate is 4 % high and the pressure at the nozzle correspondingly low, or that a partially blocked inlet has cut the flow to 11 L/min while the pump gauge, which sits upstream of a long hose, still reads 30 kPa.
Flow through a sharp-edged orifice scales with the square of the bore. Going from 6.3 to 6.38 mm raises the area by 2.6 %, so at a fixed pressure the flow rises about 2.6 %; if the regulator instead holds 12.5 L/min, the pressure at the nozzle falls and the jet loses the penetration that the standard is trying to reproduce. Two hundredths of a millimetre per year is an ordinary wear rate for a brass nozzle running filtered tap water, which puts a nozzle 0.1 mm oversize within five years.
Wear is rarely uniform either. A burr left by the drill, or a nozzle that has been cleaned with a hardened wire, develops an oval bore that deflects the stream by two or three degrees. At 2.5 m that is a lateral shift of 100 to 130 mm, enough to move the impact point off the seal you meant to test and onto the flat panel beside it. The test then passes for the wrong reason, and nobody notices until the same product fails at a different laboratory.
Catch and time is accurate enough for a bench check and needs no instrument beyond a scale. Run the jet into a tared container for 30 s, weigh, and convert. 12.5 L/min is 6.25 kg in 30 s; 100 L/min is 50 kg in 30 s, which is why IPX6 benches use a large drum or a recirculating sump with an inline meter instead. Correct for density at the water temperature, 0.998 kg/L at 20 °C and 0.992 kg/L at 40 °C, a difference of 0.6 % that sits uncomfortably close to the 5 % tolerance if the water has been recirculating all afternoon.
Measure at the pressure you actually run, read on a gauge mounted at the nozzle rather than at the pump, and do it with the nozzle in its normal position on the manifold. A gauge 3 m upstream of the nozzle across a 12 mm hose reads high by 5 to 15 kPa at these flows, which is a third of the IPX5 pressure budget.
| Parameter | IPX5 | IPX6 |
|---|---|---|
| Nozzle bore | 6.3 mm | 12.5 mm |
| Delivery rate | 12.5 L/min ± 5 % | 100 L/min ± 5 % |
| Pressure at the nozzle | 30 kPa | 100 kPa |
| Distance to enclosure | 2.5 to 3 m | 2.5 to 3 m |
| Duration | at least 3 min | at least 3 min |
| 30 s catch, at 20 °C | 6.25 kg | 50 kg |
Three triggers, any one of which is enough. Measured flow at the reference pressure outside ±5 %. Bore ovality above 0.05 mm, measured with a pin gauge or a two-axis bore check. A visible deflection of the stream when the nozzle is held against a straight edge at 1 m. Replacement is cheap; a week of retesting after a disputed result is not.
Stainless nozzles in clean water last for years and are worth the price difference. Brass nozzles in hard water scale up at the bore in months, and the scale does not simply reduce flow, it also roughens the exit and spreads the cone. If the local water is above about 200 mg/L hardness, either fit a softener on the bench supply or plan on descaling quarterly with a dilute acid and re-checking the bore afterwards, because descaling removes metal as well as scale.
Keep a single nozzle that is never used for product testing, checked annually, and use it to confirm the bench rather than the product. When a result is disputed, running the reference nozzle on the same pump, hose and gauge separates a bench problem from a product problem in about ten minutes. Without it, every argument becomes a discussion about the sample.
Five lines per nozzle, once a quarter. It is the cheapest quality record on the bench and the one that most often settles a customer complaint, because it is the only document that says what the jet actually was on the day the test ran.