IPX5 and IPX6 both specify the nozzle at 2.5 to 3 m from the enclosure. Three things about that sentence get lost in practice: it is a range rather than a value, the distance is to the nearest surface of the enclosure rather than to its centre, and it is measured from the nozzle exit rather than from the body of the gun or the end of the lance.
On a 600 mm wide cabinet, measuring to the centre instead of the nearest face adds 300 mm, which moves a nominal 2.8 m to 3.1 m and puts the test outside the range. On a small sample, measuring from the lance body rather than the nozzle exit can subtract 150 mm, which is the same error in the other direction. Neither mistake is visible in a photograph of the setup.
A free jet spreads and slows. Over the 0.5 m allowed by the range, the impact pressure on a flat surface falls by roughly 10 to 20 %, depending on the nozzle and the pressure. That is a larger effect than a nozzle worn by 0.05 mm, and it acts in the same direction: a test run at 3.1 m is a gentler test than one run at 2.5 m.
For IPX6 the effect is smaller in relative terms because the jet is more coherent at 12.5 mm, but the absolute pressure is higher and the consequences of a short distance are worse: at 2 m a 100 kPa jet can lift a gasket that would survive at 2.8 m.
| Distance from nozzle exit | Relative impact pressure | When it is used |
|---|---|---|
| 2.0 m | About 115 % of nominal | Outside the range, not a valid test |
| 2.5 m | About 105 % | Lower end of the range, harshest valid |
| 2.8 m | Nominal | Middle of the range, most reproducible |
| 3.0 m | About 92 % | Upper end of the range, mildest valid |
| 3.2 m | About 85 % | Outside the range, not a valid test |
Put a mark on the floor rather than measuring every time. Measure once from the nozzle exit, with the gun in its normal working position and the lance at its normal angle, to the nearest face of a sample positioned at its normal place on the turntable. Mark the floor, and mark the sample position. From then on the setup is a matter of putting two things on two marks.
The measurement has to be repeated whenever the sample size changes, because the nearest face moves. A 200 mm box and a 900 mm cabinet cannot share the same floor mark if both are to sit inside the range, and pretending otherwise is how a test ends up at 2.2 m without anyone noticing.
Write three numbers on the setup sheet: distance from nozzle exit to the nearest face, the height of the nozzle above the turntable, and the angle of the lance. Add a sketch with the sample outline and the two marks. Anyone who has to repeat the test a year later will have a chance of reproducing it, which is more than can be said for a sheet that says 'distance per standard'.